Particle Image Velocimetry Evaluation of Hemodynamics Proximal to the Kissing Stent Configuration in the Aorto-Iliac Bifurcation

Author:

Williamson Petra N.1ORCID,Docherty Paul D.1,Khanafer Adib2,Yazdi Sina G.1,Jermy Mark1,Kabaliuk Natalia1ORCID,Murton Benjamin1ORCID

Affiliation:

1. Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand

2. Vascular, Endovascular, & Renal Transplant Unit, Christchurch Hospital, Canterbury District Health Board, Christchurch, New Zealand

Abstract

Purpose: The kissing stent (KS) method is low-risk compared with open surgery techniques. It is often used to treat aorto-iliac occlusive disease (AIOD). Deployment of the KS geometry has a high technical success rate. However, stent patency reduces in the first 5 years potentially due to deleterious flow behavior. Potentially harmful hemodynamics due to the KS were investigated in vitro. Methodology: A compliant phantom of the aorto-iliac bifurcation was manufactured. Two surrogate stent-grafts were deployed into the phantom in the KS configuration to investigate effects of the presence of the stents, including the compliance mismatch they cause, on the hemodynamics proximal and distal to the KS. The investigation used pulsatile flow through a flow circuit to simulate abdominal aortic flow. Particle image velocimetry (PIV) was used to quantify the hemodynamics. Results: PIV identified peak proximal and distal velocity in vitro was 0.71 and 1.40m·s−1, respectively, which were within physiological ranges. Throughout systole, flow appeared normal and undisturbed. A lumen wall collapse in the sagittal plane formed during late systole and continued to early diastole proximal to the aorto-iliac bifurcation, distal to the inlet stent position. The wall collapse led to disturbed flow proximal to the stented region in early diastole producing potential recirculation zones and abnormal flow patterns. Conclusion: The normal systolic flow behavior indicates the KS configuration is unlikely to cause an inflammatory response of the arterial walls. The collapse has not been previously identified and may potentially cause long-term patency reduction. It requires further investigation. Clinical Impact The role of this article is to provide further insight into the haemodynamic behavior through a stented aorto-iliac artery. The results of this investigation will improve the understanding of the effects that using the kissing stent method may have on a patient and help to identify high risk regions that may require more detailed monitoring. This paper also develops the in vitro modelling techniques that will enable further research that cannot be carried out within patients.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3