The Effect of a Global, Subject, and Device-Specific Model on a Noninvasive Glucose Monitoring Multisensor System

Author:

Caduff Andreas1,Zanon Mattia1,Mueller Martin1,Zakharov Pavel1,Feldman Yuri2,De Feo Oscar1,Donath Marc3,Stahel Werner A.4,Talary Mark S.1

Affiliation:

1. Biovotion AG, Zurich, Switzerland

2. Department of Applied Physics, Hebrew University of Jerusalem, Jerusalem, Israel

3. Clinic for Endocrinology and Diabetes, University Hospital Basel, Basel, Switzerland

4. Seminar for Statistics, ETH Zurich, Zurich, Switzerland

Abstract

Background: We study here the influence of different patients and the influence of different devices with the same patients on the signals and modeling of data from measurements from a noninvasive Multisensor glucose monitoring system in patients with type 1 diabetes. The Multisensor includes several sensors for biophysical monitoring of skin and underlying tissue integrated on a single substrate. Method: Two Multisensors were worn simultaneously, 1 on the upper left and 1 on the upper right arm by 4 patients during 16 study visits. Glucose was administered orally to induce 2 consecutive hyperglycemic excursions. For the analysis, global (valid for a population of patients), personal (tailored to a specific patient), and device-specific multiple linear regression models were derived. Results: We find that adjustments of the model to the patients improves the performance of the glucose estimation with an MARD of 17.8% for personalized model versus a MARD of 21.1% for the global model. At the same time the effect of the measurement side is negligible. The device can equally well measure on the left or right arm. We also see that devices are equal in the linear modeling. Thus hardware calibration of the sensors is seen to be sufficient to eliminate interdevice differences in the measured signals. Conclusions: We demonstrate that the hardware of the 2 devices worn on the left and right arms are consistent yielding similar measured signals and thus glucose estimation results with a global model. The 2 devices also return similar values of glucose errors. These errors are mainly due to nonstationarities in the measured signals that are not solved by the linear model, thus suggesting for more sophisticated modeling approaches.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Bioengineering,Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 21 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. The effect of ionic redistributions on the microwave dielectric response of cytosol water upon glucose uptake;European Biophysics Journal;2024-04-22

2. On Non-Invasive Glucose Measurements;Journal of Contemporary Physics (Armenian Academy of Sciences);2022-12

3. Comparative assessment of blood glucose monitoring techniques: a review;Journal of Medical Engineering & Technology;2022-07-27

4. The inhibition of glucose uptake to erythrocytes: microwave dielectric response;European Biophysics Journal;2022-05-09

5. Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria;Journal of Diabetes Science and Technology;2021-03-09

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3