Affiliation:
1. Department of Chemical and Biological Engineering, Rensselaer Polytechnic Institute, Troy, New York
Abstract
Intensive care unit (ICU) blood glucose control algorithms were reviewed and analyzed in the context of linear systems theory and classical feedback control algorithms. Closed-loop performance was illustrated by applying the algorithms in simulation studies using an in silico model of an ICU patient. Steady-state and dynamic input-output analysis was used to provide insight about controller design and potential closed-loop performance. The proportional-integral-derivative, columnar insulin dosing (CID, Glucommander-like), and glucose regulation for intensive care patients (GRIP) algorithms were shown to have similar features and performance. The CID strategy is a time-varying proportional-only controller (no integral action), whereas the GRIP algorithm is a nonlinear controller with integral action. A minor modification to the GRIP algorithm was suggested to improve the closed-loop performance. Recommendations were made to guide control theorists on important ICU control topics worthy of further study.
Subject
Biomedical Engineering,Bioengineering,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
18 articles.
订阅此论文施引文献