In an article in the Journal of Diabetes Science and Technology, Macleod and coworkers describe an evaluation of LifeScan glucose meters that focus on the effects of sample types and comparison methods. They make a valid point that these factors influence the accuracy observed in evaluations and recommend the comparison method be the one recommended by the manufacturer for the sample type in the intended use statement. Yet, the recommended comparison method is not a reference method. The accuracy hierarchy of definitive, reference, and field methods originally described by Tietz should remind one that virtually all glucose meter evaluations use commercially available field methods as the comparison method. Finally, one should not neglect the FDA adverse event database as a way to assess glucose meter performance.