Accuracy of the SEVEN® Continuous Glucose Monitoring System: Comparison with Frequently Sampled Venous Glucose Measurements

Author:

Zisser Howard C.1,Bailey Timothy S.2,Schwartz Sherwyn3,Ratner Robert E.4,Wise Jonathan5

Affiliation:

1. Sansum Diabetes Research Institute and University of California at Santa Barbara, Santa Barbara, California

2. Advanced Metabolic Care + Research, Escondido, California

3. Diabetes and Glandular Disease Clinic, San Antonio, Texas

4. MedStar Research Institute, Washington, D.C.

5. Diabetes and Metabolism Associates, Metairie, Louisiana

Abstract

Background: The purpose of this study was to compare the accuracy of measurements obtained from the DexCom™ SEVEN® system with Yellow Springs Instrument (YSI) laboratory measurements of venous blood glucose. Methods: Seventy-two subjects with insulin-requiring diabetes, aged 18–71, were enrolled in a multicenter, prospective single-arm study. All participants wore the SEVEN continuous glucose monitoring (CGM) system for one, 7-day wear period. Calibration with capillary finger stick measurements was performed 2 hours after sensor insertion and once every 12 hours thereafter. A subset of subjects (28) wore two systems simultaneously to assess precision. All subjects participated in one, 10-hour in-clinic session on day 1, 4, or 7 of the study to compare CGM measurements against a laboratory method (YSI analyzer) using venous measurements taken once every 20 minutes. Carbohydrate consumption and insulin dosing were adjusted in order to obtain a broad range of glucose values. Results: Comparison of CGM measurements with the laboratory reference method ( n = 2318) gave mean and median absolute relative differences (ARDs) of 16.7 and 13.2%, respectively. The percentage was 70.4% in the clinically accurate Clarke error grid A zone and 27.5% in the benign error B zone. Performance of the SEVEN system was consistent over time with mean and median ARD lowest on day 7 as compared to YSI (13.3 and 10.2%, respectively). Average sensor time lag was 5 minutes. Conclusions: Measurements of the DexCom SEVEN system were found to be consistent and accurate compared with venous measurements made using a laboratory reference method over 7 days of wear.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Bioengineering,Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 67 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3