Redundancy in Glucose Sensing

Author:

Sharifi Amin12,Varsavsky Andrea3,Ulloa Johanna3,Horsburgh Jodie C.2,McAuley Sybil A.12,Krishnamurthy Balasubramanian1,Jenkins Alicia J.124,Colman Peter G.5,Ward Glenn M.1,MacIsaac Richard J.12,Shah Rajiv3,O’Neal David N.12

Affiliation:

1. Department of Endocrinology & Diabetes, St Vincent’s Hospital Melbourne, Melbourne, Australia

2. Department of Medicine, St Vincent’s Hospital, University of Melbourne, Melbourne, Australia

3. Sensor R & D, Medtronic Diabetes, Northridge, CA, USA

4. NHMRC Clinical Trials Centre, Sydney, Australia

5. Department of Diabetes and Endocrinology, Royal Melbourne Hospital, Melbourne, Australia

Abstract

Background: Current electrochemical glucose sensors use a single electrode. Multiple electrodes (redundancy) may enhance sensor performance. We evaluated an electrochemical redundant sensor (ERS) incorporating two working electrodes (WE1 and WE2) onto a single subcutaneous insertion platform with a processing algorithm providing a single real-time continuous glucose measure. Methods: Twenty-three adults with type 1 diabetes each wore two ERSs concurrently for 168 hours. Post-insertion a frequent sampling test (FST) was performed with ERS benchmarked against a glucose meter (Bayer Contour Link). Day 4 and 7 FSTs were performed with a standard meal and venous blood collected for reference glucose measurements (YSI and meter). Between visits, ERS was worn with capillary blood glucose testing ≥8 times/day. Sensor glucose data were processed prospectively. Results: Mean absolute relative deviation (MARD) for ERS day 1-7 (3,297 paired points with glucose meter) was (mean [SD]) 10.1 [11.5]% versus 11.4 [11.9]% for WE1 and 12.0 [11.9]% for WE2; P < .0001. ERS Clarke A and A+B were 90.2% and 99.8%, respectively. ERS day 4 plus day 7 MARD (1,237 pairs with YSI) was 9.4 [9.5]% versus 9.6 [9.7]% for WE1 and 9.9 [9.7]% for WE2; P = ns. ERS day 1-7 precision absolute relative deviation (PARD) was 9.9 [3.6]% versus 11.5 [6.2]% for WE1 and 10.1 [4.4]% for WE2; P = ns. ERS sensor display time was 97.8 [6.0]% versus 91.0 [22.3]% for WE1 and 94.1 [14.3]% for WE2; P < .05. Conclusions: Electrochemical redundancy enhances glucose sensor accuracy and display time compared with each individual sensing element alone. ERS performance compares favorably with ‘best-in-class’ of non-redundant sensors.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Bioengineering,Endocrinology, Diabetes and Metabolism,Internal Medicine

Cited by 13 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Clinical Performance Evaluation of Continuous Glucose Monitoring Systems: A Scoping Review and Recommendations for Reporting;Journal of Diabetes Science and Technology;2023-08-20

2. Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria;Journal of Diabetes Science and Technology;2021-03-09

3. Assessing the Accuracy of Smart Stents for Monitoring Coronary Artery Restenosis;2020 IEEE-EMBS Conference on Biomedical Engineering and Sciences (IECBES);2021-03-01

4. Wearable multiplexed biosensor system toward continuous monitoring of metabolites;Biosensors and Bioelectronics;2020-04

5. Human-in-the-Loop Insulin Dosing;Journal of Diabetes Science and Technology;2019-12-04

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3