Affiliation:
1. Laboratory Medicine, Division of Clinical Biochemistry, Saskatchewan Health Authority, Saskatoon, Saskatchewan, Canada
Abstract
Background: Our objective was to determine the effect of therapeutic concentrations of N-acetylcysteine, following intravenous infusion, on the measurement of blood glucose using a Roche Diagnostics glucose dehydrogenase-linked glucose meter compared to hospital laboratory methods. Methods: N-acetylcysteine was added to aliquots of blood, with glucose promptly measured by the glucose meter, blood gas analyzer (glucose oxidase comparative method) and following centrifugation, plasma glucose measured with a hexokinase spectrophotometric comparative method. Glucose results were evaluated with linear regression and Bland Altman plots. Results: In the presence of NAC, at concentrations greater than 5 mg/dL (0.31 mmol/L), positively biased glucose meter results were compared to the clinical laboratory results. Multivariate linear regression revealed that NAC-mediated meter results are influenced by NAC and glucose concentrations. Conclusions: The addition of therapeutic concentrations of NAC to blood produces statistically significant positive biases when measured with the glucose dehydrogenase linked glucose meter device.
Subject
Biomedical Engineering,Bioengineering,Endocrinology, Diabetes and Metabolism,Internal Medicine
Cited by
6 articles.
订阅此论文施引文献