Use of O-arm with neuronavigation in percutaneous vertebroplasty reduces the surgeon’s exposure to intraoperative radiation

Author:

Wojdyn Maciej1,Pierzak Olaf1,Zapałowicz Krzysztof2,Radek Maciej1

Affiliation:

1. Department of Neurosurgery, Surgery of Spine and Peripheral Nerves, Medical University of Lodz, University Hospital WAM-CSW, Lodz, Poland

2. Department of Neurosurgery, Prof. Gibiński Central Clinical Hospital, Katowice, Poland

Abstract

IntroductionPercutaneous vertebroplasty is commonly used to treat spinal fractures. The authors compare radiation exposure as potential risk for the surgical team during vertebroplasty guided by O-arm combined with neuronavigation versus vertebroplasty guided by C-arm fluoroscopy.Material and methodsThe clinical material consisted of a group of 29 patients (44 vertebrae) with fractures of the thoracolumbar spine treated with percutaneous vertebroplasty guided by O-arm with neuronavigation. In this new method, the operating room staff leaves the operating room for the duration of the 3D scan of the appropriate spine section using the O-arm. In the next stage, the needle of the vertebroplasty system is introduced using only neuronavigation without the need for a radiological view. Finally, the cement injection was made under O-arm fluoroscopic control. The comparison group consisted of a group of 35 patients (40 vertebrae) treated with the classical method using C-arm fluoroscopy. The two methods were compared in terms of the average dose of emitted ionizing radiation through the device (O-arm vs. C-arm) to which surgeons are exposed during percutaneous vertebroplasty.ResultsAs a result of vertebroplasty procedures guided by neuronavigation, a statistically significant difference between the values of mean dose of radiation emitted by O-arm and C-arm systems was noted. The O-arm emitted 912 cGy/cm2 vs. 1722 cGy/cm2 emitted by the C-arm during fluoroscopically assisted procedures and 601.28 cGy/cm2 vs. 1506.86 cGy/cm2 per vertebrae.ConclusionsDuring vertebroplasty with the O-arm combined with neuronavigation the radiation dose is significantly lower as compared with the C-arm used for fluoroscopic guidance, minimizing the potential risk of radiation exposure to surgeons.

Publisher

Termedia Sp. z.o.o.

Subject

General Medicine

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3