Dosimetric verification of radiotherapy treatment planning system at TMSS Cancer Center, Bogura, Bangladesh

Author:

Mehrab Hassan Udoy ,Rajada Khatun ,Mokhlesur Rahman ,Motiur Rahman ,Fajle Rabby ,Shirin Akter

Abstract

Authentication of TPS is the component of the QA program for RTPS and planning process. So, verification of dose calculation algorithm is essential to measure the efficiency of the dose calculations. The goal is to assess how far the dosage measured deviates from the dose predicted by the treatment planning system (TPS) algorithm using a CIRS Thorax phantom. It has been done to measure the dosimetric variations of 6MV, 10MV and 15MV photon beams. The phantom was initially scanned using a CT-simulator, and the raw data from the gated scans were then exported to TPS to perform the planning. Using this planning, the phantom was irradiated for four different cases. The computed dose was then compared with the measured dose to ensure TPS validity. The findings demonstrate that the variation between the measured doses in altered holes of the CIRS thorax phantom and the calculated doses by TPS was within the allowed limit except few points of four cases. Max. and Min. deviations for 6MV were 3.87% and 0.05% respectively. On the other hand, Max. and Min. deviations for 10MV were 7.87% and 0.12% respectively and that for 15MV are 4.44% and 0.02% respectively. Satisfactory results were obtained for lung and bone substitute. The results demonstrate the accuracy of the doses determined by the treatment planning system algorithm using a thorax phantom, and they also demonstrate the competency of using the phantom for regular verification. In case of external beam radiation therapy with AAA algorithm, the CIRS Thorax Phantom is suitable for dose computation in heterogeneous medium. However it is more compatible for 6MV than 10MV and 15MV.

Publisher

GSC Online Press

Subject

General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3