Affiliation:
1. Alturas Analytics, Inc, 1324 Alturas Drive, Moscow, ID 83843, USA
Abstract
Background: Dried matrix spot techniques were employed to validate an HPLC–MS/MS assay for the determination of dexamethasone in clear Yorkshire pig synovial fluid using 15 µl of sample. We have adopted the term dried matrix spot to indicate that the techniques used for dried blood spots can be applied to nonblood matrices. The dried matrix spot method employs a color-indicating process developed at Alturas Analytics that enhances the ability to analyze transparent fluids spotted onto collection paper by allowing the analyst to visually verify the location of the dried sample spot. Results: The method was shown to be accurate (±4.3%) and precise (14.2% at the LLOQ and ≤10.0% at all other concentrations) across the dynamic range of the assay. Conclusion: The method shows the potential application of dried matrix spot techniques for the analysis of transparent biological fluids.
Subject
Medical Laboratory Technology,Clinical Biochemistry,General Pharmacology, Toxicology and Pharmaceutics,General Medicine,Analytical Chemistry
Cited by
21 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献