Author:
Lazurik V.T.,Lazurik V.M.,Popov G.Ph.,Zimek Z.
Abstract
A new conservative computational procedure is proposed for calculating the average absorbed energy of electron radiation in objects undergoing radiation processing. Semi-empirical relationships were obtained and software for dosimetry of electron radiation was developed. A comparison was made of the results obtained using the proposed method with the results of known and traditional methods of electron radiation dosimetry as well as with the results of measurements performed at the sterilization center of the Institute of Nuclear Chemistry and Technology, Warsaw, Poland was carried out.
Publisher
Problems of Atomic Science and Technology
Reference9 articles.
1. Z. Zimek, L. Waliś, A.G. Chmielewski. Industrial facility for radiation sterilization of medical devices // Radiat. Phys. Chem. 1993, v. 42, p. 571-572.
2. Z.P. Zagórski. Dependence of dept-dose distribution on the energy spectrum of 5 to 13 MeV electron beams // Radiat. Phys. Chem. 1983, v. 22, № 3-5, p. 409-418.
3. V.T. Lazurik, V.M. Lazurik, G. Popov, Z. Zimek. Determination of electron beam parameters on radiation-technological facility for simulation of radiation processing // East European Journal of Physics. 2014, v. 1, № 3, p. 74-78.
4. V.M. Lazurik, V.T. Lazurik, G. Popov, Z. Zimek. Two-parametric model of electron beam in computational dosimetry for radiation processing // Radiat. Phys. Chem. 2016, v. 124, p. 230-234.
5. V.T. Lazurik, V.M. Lazurik, G.Ph. Popov, Z. Zimek. Method of Dosimetry Based on a Two-Parametric Model of Electrons Beam for Radiation Processing // Problems of Atomic Science and Technology. 2017, № 6, p. 137-141.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献