Affiliation:
1. From the Department of Anaesthesiology, Unidade Local de Saúde da Arrábida, E.P.E., Hospital São Bernardo, Setúbal, Portugal.
Abstract
Incorrect bispectral index (BIS) values have been reported due to interference with this monitoring system. We report a case of a 46-year-old woman who underwent liposuction and breast lipofilling, where we observed a misinterpretation by the BIS algorithm that has not yet been reported. Concurrently with abdominal and thigh liposuction, an increase in the BIS value was observed. The importance of examining electroencephalogram (EEG) and density spectral array (DSA) readings during liposuction procedures is highlighted in this case report, extending our observations beyond just the numerical BIS value, which is not always reliable.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Reference15 articles.
1. A primer for EEG signal processing in anesthesia.;Rampil;Anesthesiology,1998
2. Bispectral index monitoring and intraoperative awareness.;Chang;ATOTW WFSA,2019
3. Bispectral index monitoring prevent awareness during total intravenous anesthesia: a prospective, randomized, double-blinded, multi-center controlled trial.;Zhang;Chin Med J (Engl),2011
4. Bispectral index for improving anesthetic delivery and post-operative recovery.;Punjasawadwong;Cochrane Database Sys Rev,2014
5. Effects of bispectral index monitoring on ambulatory anesthesia: a meta-analysis of randomized controlled trials and cost analysis.;Liu;Anesthesiology,2004
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献