Abstract
Background: To propose a new term (‘construct efficiency’) for the evaluation of multi strands flexor tendon repairs using different suture materials. Methods: A total of twenty specimens from 4-0 braided polyblend sutures (FiberLoop/FiberWire; Arthrex, Naples, FL) and 4-0 nylon sutures (Supramid Extra II; S. Jackson, Inc., Alexandria, VA) were subjected to tensile testing using Pneumatic Cord-and-Yarn Grips (Instron Corp., Canton MA, USA). The ultimate tensile strengths of the suture materials were measured. The expected repair strengths and construct efficiencies were computed based on the experimental results and from available literature on actual repair strengths of the 4-strand Becker, Cruciate repairs and 6-strand Tang, modified Lim-Tsai repairs. Results: The ultimate tensile strength of nylon suture was 15.4 ± 0.6N, lower than that of braided polyblend suture (45.3 ± 2.3N) with a difference of 194%. The construct efficiency of multi strand repairs varied with respect to different repair techniques and suture materials. It was found that the Becker repairs using FiberWire had the highest construct efficiency (55.7%) followed by the modified Lim-Tsai using Supramid (50.9%), Tang repair using Supramid (49.8%), Cruciate repair using Fiberwire (49.1%), and modified Lim-Tsai repair using FiberLoop (33.5%). Conclusions: The construct efficiency is more accurate in showing that, in terms of biomechanical strength, the use of FiberWire for the 4-strand Becker and Cruciate repair is more efficient than that of using FiberLoop for 6-strand modified Lim-Tsai repair.
Publisher
World Scientific Pub Co Pte Lt
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献