Study of Reliability and Validity of the Load Cell-Type Hand Dynamometer Compared to the Jamar Dynamometer and the Number of Reliable Grip Strength Measurements

Author:

KANAUCHI Yumiko1,MURASE Tsuyoshi2,NISHIWAKI Masao3,ODAGIRI Masahiro4,WANEZAKI Yoshihiro5,ISHIKAWA Hitoshi6

Affiliation:

1. Department of Rehabilitation, Yamagata City Hospital Saiseikan, Yamagata, Japan

2. Department of Orthopaedic Surgery, Bell Land General Hospital, Osaka, Japan

3. Hand and Elbow Surgery Center, Department of Orthopaedic Surgery, Kawasaki Municipal Hospital, Kawasaki, Japan

4. Niigata Hand Surgery Foundation, Niigata, Japan

5. Department of Orthopedic Surgery, Yamagata University Faculty of Medicine, Yamagata, Japan

6. Department of Physical Therapy, Yamagata Prefectural University of Health Science, Yamagata, Japan

Abstract

Background: Standardised measurement protocols for grip strength remained unclear due to variations in values depending on the device and measurement method. The load cell hand dynamometer has recently been developed. This study aims to investigate the reliability of the load cell dynamometer by comparing it to the Jamar dynamometer, which is considered the gold standard, and to identify a reliable and practical measurement method. Methods: This study included 80 healthy hospital workers (mean age of 40.1 years). All measurements were performed seated, with the elbow flexed 90° and the grip span at the second handle (approximately 50 mm) for the Jamar dynamometer, and with the elbow extended and the grip span fixed at 55 mm for the load cell dynamometer. Grip strength was measured three times on each hand using two dynamometers, and the same tests were repeated on different days. Test–retest reliability, the association between the two devices and the agreement between the two measurement methods were assessed using the intraclass correlation coefficient (ICC), Pearson correlation and the Bland–Altman analysis. Results: The ICC of the one measurement was lower than that of three measurements for both dynamometers, but was above 0.858 in all groups, indicating sufficient reliability with one-time measurement. Additionally, the ICC for different days revealed good reliability (Jamar: >0.830, load cell: >0.772). The load cell dynamometer showed significantly lower values in all measurements despite the excellent correlation (r > 0.70) and the agreement between the two dynamometers. Conclusions: This study revealed sufficient reliability of the load cell dynamometer with the standardised measurement method, but it should be noted the lower values compared to the Jamar dynamometer. Additionally, one-time measurement reliability is adequate for practical use by standardising the measurement methods for both dynamometers. Level of Evidence: Level III (Diagnostic)

Publisher

World Scientific Pub Co Pte Ltd

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3