Affiliation:
1. School of Automation Science and Electrical Engineering, Beihang University, Beijing 100191, P. R. China
Abstract
To improve the efficiency and safety of secretion clearance, a novel ventilator (SC ventilator) with an automatic secretion clearance function is proposed. To lay a foundation for the optimization of the SC ventilator, the basic mathematical model of the ventilation system is derived. By selecting the appropriate reference values, the basic mathematical model is transformed to a dimensionless expression for simulation. Through the experimental and simulation study on the SC ventilation system, it can be concluded that: firstly, the mathematical model is proved to be authentic and reliable. Secondly, the influences of the three key parameters on the dynamics of the SC ventilation system are carried out. Last, to guarantee the pressure in the lung is higher than the expiratory positive airway pressure, the dimensionless minimum pressure in the flexible tube should be set higher than 0.9164, the dimensionless suction pressure should be set higher than 0.79.
Publisher
World Scientific Pub Co Pte Lt
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献