Abstract
AbstractEvery year biotechnology labs generate a combined total of ∼5.5 million tons of plastic waste. As the global bioeconomy expands, biofoundries will inevitably increase plastic consumption in-step with synthetic biology scaling. Decontamination and reuse of single-use plastics could increase sustainability and reduce recurring costs of biological research. However, throughput and variable cleaning quality make manual decontamination impractical in most instances. Automating single-use plastic cleaning with liquid handling robots makes decontamination more practical by offering higher throughput and consistent cleaning quality. However, open-source, validated protocols using low-cost lab robotics for effective decontamination of plasticware—facilitating safe reuse—have not yet been developed. Here we introduce and validate TidyTron: a library of protocols for cleaning micropipette tips and microtiter plates that are contaminated with DNA,E. coli, andS. cerevisiae. We tested a variety of cleaning solutions, contact times, and agitation methods with the aim of minimizing time and cost, while maximizing cleaning stringency and sustainability. We tested and validated these cleaning procedures by comparing fresh versus cleaned tips and plates for contamination with cells, DNA, or cleaning solutions. We assessed contamination by measuring colony forming units by plating, PCR efficiency and DNA concentration by qPCR, and event counts and debris by flow cytometry. Open source cleaning protocols are available athttps://github.com/PlantSynBioLab/TidyTronand hosted on a graphical user interface athttps://jbryantvt.github.io/TidyTron/.
Publisher
Cold Spring Harbor Laboratory