Abstract
AbstractPotency assessment of monoclonal antibodies or corresponding biosimilars in cell-based assays is an essential prerequisite in biopharmaceutical research and development. However, cellular bioassays are still subject to limitations in sample throughput, speed, and often need costly reagents or labels as they are based on an indirect readout by luminescence or fluorescence. In contrast, whole-cell Matrix-Assisted Laser Desorption/Ionization Time-of-Flight (MALDI-TOF) Mass Spectrometry (MS) has emerged as a direct, fast and label-free technology for functional drug screening being able to unravel the molecular complexity of cellular response to pharmaceutical reagents. However, this approach has not yet been used for cellular testing of biologicals. In this study, we have conceived, developed and bench-marked a label-free MALDI-MS based cell bioassay workflow for the functional assessment of complement-dependent cytotoxicity (CDC) of Rituximab antibody. By computational evaluation of response profiles followed by subsequentm/zfeature annotation via fragmentation analysis and trapped ion mobility MS, we identified adenosine triphosphate and glutathione as readily MS-assessable metabolite markers for CDC and demonstrate that robust concentration-response characteristics can be obtained by MALDI-TOF MS. Statistical assay performance indicators suggest that whole-cell MALDI-TOF MS could complement the toolbox for functional cellular testing of biopharmaceuticals.
Publisher
Cold Spring Harbor Laboratory