Abstract
ABSTRACTAn automated ultra-microtome capable of sectioning thousands of ultrathin sections onto standard TEM slot grids was developed and used to section: a completeDrosophila melanogasterfirst-instar larva, three sections per grid, into 4,866 34-nm-thick sections with a cutting and pickup success rate of 99.74%; 30 microns of mouse cortex measuring roughly 400 um x 2000 um at 40 nm per slice; and a full adultDrosophilabrain and ventral nerve column into 9,300 sections with a pickup success rate of 99.95%. The apparatus uses optical interferometers to monitor a reference distance between the cutting knife and multiple sample blocks. Cut sections are picked up from the knife-boat water surface while they are still anchored to the cutting knife. Blocks without embedded tissue are used to displace tissue-containing sections away from the knife edge so that the tissue regions end up in the grid slot instead of on the grid rim.
Publisher
Cold Spring Harbor Laboratory