Abstract
Here we describe some practical concerns surrounding the scanning of microarray slides that have been hybridized with fluorescent dyes. We use a laser scanner that has two lasers, each set to excite a different fluor, and separate detectors to capture emission from each fluor. The laser passes over an address (position on the scanned surface) and the detectors capture photons emitted from each address. Two superimposed image files are written that carry intensities for each channel for each pixel of the image scan. These are the raw data. Image analysis software is used to identify and summarize the intensities of the pixels that make up each spot. After comparison to background pixels, the processed intensity levels representing the gene expression measurements are associated with the identity of each spot.