Abstract
AbstractBackgroundCardiovascular implantable electronic devices (CIEDs) such as permanent pacemakers, implantable cardioverter-defibrillators, and cardiac resynchronization therapy devices alleviate morbidity and mortality in various diseases. There is a paucity of real-world data on CIED complications and trends.ObjectivesDescribe trends in noninfectious CIED complications over the past three decades in Olmsted County.MethodsThe Rochester Epidemiology Project is a medical records linkage system comprising records of over 500,000 residents of Olmsted County from 1966-current. CIED implants between 1988-2018 were determined. Trends in noninfectious complications within 30 days of implant were analyzed.Results175 out of 2536 (6.9%) patients who received CIED experienced device complications. 3.8% of the implants had major complications requiring intervention. Lead dislodgement was the most common (2.9%), followed by hematoma (2.1%). Complications went up from 1988 to 2005, then showed a downtrend until 2018, driven by a decline in hematomas in the last decade (p<0.01). Those with complications were more likely to have prosthetic valves. Obesity appeared to have a protective effect in a multivariate regression model. The mean Charlson comorbidity score has trended up over the 30 years.ConclusionsOur study describes a real-world trend of CIED complications over three decades. Lead dislodgements and hematomas were the most common complications. Complications have declined over the last decade due to safer practices and a better understanding of anticoagulant management.
Publisher