Design, development, and preliminary assessment of a novel peripheral intravenous catheter aimed at reducing early failure rates

Author:

Doyle Barry J.ORCID,Kelsey Lachlan J.,Shelverton Caroline,Abbate Gabriella,Ainola Carmen,Sato Noriko,Livingstone Samantha,Bouquet Mahe,Passmore Margaret R,Wilson Emily S.,Colombo Sebastiano,Sato Kei,Liu Keibun,Heinsar Silver,Wildi Karin,Carr Peter J.,Suen Jacky,Fraser John,Li Bassi Gianluigi,Keogh Samantha

Abstract

ABSTRACTBackgroundPeripheral intravenous catheters (PIVCs) are the most commonly used invasive medical device, yet despite best efforts by end-users, PIVCs experience unacceptably high early failure rates. We aimed to design a new PIVC that reduces the early failure rate of in-dwelling PIVCs and we conducted preliminary tests to assess its efficacy and safety in a large animal model of intravenous access.MethodsWe used computer-aided design and simulation to create a PIVC with a ramped tip geometry, which directs the infused fluid away from the vein wall; we called the design the FloRamp™. We created FloRamp prototypes (test device) and tested them against a market-leading device (BD Insyte™; control device) in a highly-controlled setting with five insertion sites per device in four pigs. We measured resistance to infusion and visual infusion phlebitis (VIP) every six hours and terminated the experiment at 48 hours. Veins were harvested for histology and seven pathological markers were assessed.ResultsComputer simulations showed that the optimum FloRamp tip reduced maximum endothelial shear stress by 60%, from 12.7Pa to 5.1Pa, compared to a typical PIVC tip, and improved the infusion dynamics of saline in the blood stream. In the animal study, we found that 2/5 of the control devices were occluded after 24 hours, whereas all test devices remained patent and functional. The FloRamp created less resistance to infusion (0.73±0.81 vs 0.47±0.50, p=0.06) and lower VIP scores (0.60±0.93 vs 0.31±0.70, p=0.09) that the control device, although neither findings were significantly different. Histopathology revealed that 5/7 of the assessed markers were lower in veins with the FloRamp.ConclusionsAs PIVCs are used in almost every hospitalized patient, there is an urgent need to reduce failure rates. Herein we report preliminary assessment of a novel PIVC design, which could be advantageous in clinical settings through decreased device occlusion.

Publisher

Cold Spring Harbor Laboratory

Reference28 articles.

1. Accepted but Unacceptable

2. How common are indwelling devices in hospitalized adults? A contemporary point prevalence study in a tertiary care hospital;American Journal of Infection Control,2021

3. Burnaby B. Global market overview for vascular acceess devices and accessories 2012–2022. 2017. Contract No.: 4339819.

4. Verified Market Research. Global Peripheral IV Catheter Market: Market Size, Status and Forecast to 2028. 2020.

5. Observational Study of Peripheral Intravenous Catheter Outcomes in Adult Hospitalized Patients: A Multivariable Analysis of Peripheral Intravenous Catheter Failure;J Hosp Med,2018

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3