Author:
Ravi Nirmal,Vithyananthan Mathura,Saidu Aisha Amira
Abstract
AbstractInfrared thermometry has certain advantages over traditional oral thermometry including quick, non-invasive administration and an absence of required consumables. This study compared the performance of tympanic, temporal artery and forehead contactless thermometers with traditional oral electronic thermometer in measuring temperature in outpatients in a Nigerian secondary care hospital. A convenience sample of 100 male and 100 female adult patients (Mean age= 38.46 years, SD= 16.33 years) were recruited from a secondary care hospital in Kano, Nigeria. Temperature measurements were taken from each patient using the tympanic, temporal artery and contactless thermometers and oral electronic thermometer. Data was analyzed to assess bias and limits using scatterplots and Bland-Altman charts while sensitivity analysis was done using ROC curves. The tympanic and temporal artery thermometers systematically gave higher temperature readings compared to the oral electronic thermometer. Contactless thermometer gave lower readings compared to the oral electronic thermometer. Temporal artery thermometer had the highest sensitivity (88%) and specificity (88%) among the three infrared thermometers. Contactless thermometer showed a low sensitivity of 13% to detect fever greater than 38°C. Our study shows that replacing oral thermometers with infrared thermometers must be done with caution despite the associated convenience and cost savings.
Publisher
Cold Spring Harbor Laboratory
Reference28 articles.
1. Tympanic, infrared skin and temporal artery scan thermometers compared with rectal measurement in children: A real-life assessment;Cur. Therap. Res,2014
2. Accuracy and precision of four common peripheral temperature measurement methods in intensive care patients. Med Devices;Evid. and Res,2016
3. Temperature measurement in pediatrics: A comparison of the rectal methods versus the temporal artery method;J. Perianesth.Nurs,2012
4. Agreement between temporal artery, oral and axillary temperature measurements in the perioperative period;J. Perianesth.Nurs,2011
5. Berkosy, E. A. , O. Bag , S. Yazici , and T. Celik . Use of noncontact infrared thermography to measure temperature in children in a triage room. Med. 97(5), 2018.