Sharing ventilators in the Covid-19 pandemics. A bench study

Author:

Guérin ClaudeORCID,Cour Martin,Stevic Neven,Degivry Florian,L’Her Erwan,Louis Bruno,Argaud Laurent

Abstract

AbstractCOVID-19 pandemics sets the healthcare system to a shortage of ventilators. We aimed at assessing tidal volume (VT) delivery and air recirculation during expiration when one ventilator is divided into 2 patients. The study was performed in a research laboratory in a medical ICU of a University hospital. An ICU-dedicated (V500) and a lower-level ventilator (Elisée 350) were attached to two test-lungs (QuickLung) through a dedicated flow-splitter. A 50 mL/cmH2O Compliance (C) and 5 cmH2O/L/s Resistance (R) were set in both A and B lungs (step1), C50R20 in A / C20R20 in B (step 2), C20R20 in A / C10R20 in B (step 3), and C50R20 in A / C20R5 in B (step 4). Each ventilator was set in volume and pressure control mode to deliver 0.8L VT. We assessed VT from a pneumotachograph placed immediately before each lung, rebreathed volume, and expiratory resistance (circuit and valve). Values are median (1st-3rd quartiles) and compared between ventilators by non-parametric tests. Between Elisée 350 and V500 in volume control VT in A/B patients were 0.381/0.387 vs. 0.412/0.433L in step 1, 0.501/0.270 vs. 0.492/0.370L in step 2, 0.509/0.237 vs. 0.496/0.332L in step 3, and 0.496/0.281 vs. 0.480/0.329L in step 4. In pressure control the corresponding values were 0.373/0.336 vs. 0.430/0.414L, 0.416/0.185/0.322/0.234L, 0.193/0.108 vs. 0.176/0.092L and 0.422/0.201 vs. 0.481/0.329L, respectively (P<0.001 between ventilators at each step for each volume). Rebreathed air volume ranged between 0.7 to 37.8 ml and negatively correlated with expiratory resistance in steps 2 and 3. The lower-level ventilator performed closely to the ICU-dedicated ventilator. Due to dependence of VT to C pressure control should be used to maintain adequate VT at least in one patient when C and/or R changes abruptly and monitoring of VT should be done carefully. Increasing expiratory resistance should reduce rebreathed volume.

Publisher

Cold Spring Harbor Laboratory

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3