Abstract
AbstractPurposeRetinal oximetry is a technique based on spectrophotometry where images are analyzed with software capable of calculating vessel oxygen saturation and vessel diameter. In this study, the effect of automation of measurements of retinal vessel oxygen saturation and vessel diameter is explored.MethodsUntil now, operators have had to choose each vessel segment to be measured explicitly. A new, automatic version of the software automatically selects the vessels once the operator defines a measurement area.Five operators analyzed image pairs from the right eye of 23 healthy subjects with semiautomated retinal oximetry analysis software, Oxymap Analyzer (v2.5.1), and an automated version (v3.0). Inter- and intra-operator variability was investigated using the intraclass correlation coefficient (ICC) between oxygen saturation measurements of vessel segments in the same area of the retina.ResultsFor manual saturation measurements, the inter-rater ICC was 0.80 for arterioles and venules. For automated saturation measurements, the inter-rater ICC was 0.97 for arterioles and 0.96 for venules. For manual diameter measurements, the inter-rater ICC was 0.71 for arterioles and venules. For automated diameter measurements the inter-rater ICC was 0.97 for arterioles and 0.95 for venules. The inter-rater ICCs were different (p < 0.01) between the semiautomated and automated version in all instances.ConclusionAutomated measurements of retinal oximetry values are more repeatable compared to measurements where vessels are selected manually.