Real time monitoring of transtibial elevated vacuum prostheses: a case series on socket air pressure

Author:

Schoepp Katherine R.,Schofield Jonathon S.,Home David,Dawson Michael R.ORCID,Lou Edmond,Keri McNiel,Marasco Paul D.,Hebert Jacqueline S.ORCID

Abstract

AbstractElevated vacuum is a prosthetic suspension method used to reduce slippage between the prosthetic socket and the residual limb. Evaluation of the effectiveness of these systems is limited due to a lack of correlation to actual socket air pressure, particularly during unconstrained movements. This may explain some of the variability in functional outcomes reported in the literature. We developed a light-weight portable socket measurement system to quantify internal socket air pressure, temperature, and acceleration. We implemented the system onto the sockets of three transtibial prosthesis users with mechanical elevated vacuum pumps. Participants completed five functional tasks with and without the vacuum pumps actively connected, including the 2-Minute Walk test, 5-Times Sit-to-Stand test, 4-Square Step test, L-Test, and Figure-8 test. Results demonstrated that the use of elevated vacuum pumps produced different gait profiles and pressure ranges for each user, with significant differences between pump conditions. Two of the participants demonstrated substantially lower air pressure (higher vacuum) over time while the pump was active compared to inactive. The minimum air pressure measured at the completion of the 2-Minute Walk test was −34.6 ± 7.7 kPa, which is not as low as pressures reported in literature during benchtop experiments. One participant did not show substantial changes in pressure over time for either pump condition. Functional task performance was not significantly different between pump conditions. Correlation with accelerometer readings allowed air pressure data to be aligned with the gait cycle; peak positive pressures occurred just following initial contact of the foot in early stance, and the most negative pressures (vacuum) were observed throughout swing. This study has demonstrated the use of a portable data logging tool that may serve the clinical and research communities to quantify the operation of elevated vacuum systems, and better understand the variability of mechanical pump operation and overall system performance.

Publisher

Cold Spring Harbor Laboratory

Reference44 articles.

1. Estimating the Prevalence of Limb Loss in the United States: 2005 to 2050

2. State-of-the-art research in lower-limb prosthetic biomechanics-socket interface: A review;J Rehabil Res Dev,2001

3. Issues of importance reported by persons with lower limb amputations and prostheses;J Rehabil Res Dev,1999

4. Levy W . Chapter 26: Skin Problems of the Amputee. In: Bowrker JH , Michael JW , editors. Atlas of Prosthetics: Surgical, Prosthetic, and Rehabilitation Principles: Mosby Year Book; 1992.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3