Artifacts can be deceiving: The actual location of deep brain stimulation electrodes differs from the artifact seen on magnetic resonance images

Author:

Nuzov Noa B.,Bhusal Bhumi,Henry Kaylee R.,Jiang Fuchang,Vu Jasmine,Rosenow Joshua M.,Pilitsis Julie G.,Elahi Behzad,Golestanirad Laleh

Abstract

AbstractIntroductionDeep brain stimulation (DBS) is a common treatment for a variety of neurological and psychiatric disorders. Recent studies have highlighted the role of neuroimaging in localizing the position of electrode contacts relative to target brain areas in order to optimize DBS programming. Among different imaging methods, postoperative magnetic resonance imaging (MRI) has been widely used for DBS electrode localization; however, the geometrical distortion induced by the lead limits its accuracy. In this work, we investigated to what degree the difference between the actual location of the lead’s tip and the location of the tip estimated from the MRI artifact varies depending on the MRI sequence parameters such as acquisition plane and phase encoding direction, as well as the lead’s extracranial configuration. Accordingly, an imaging technique to increase the accuracy of lead localization was devised and discussed.MethodsWe designed and constructed an anthropomorphic phantom with an implanted DBS system following 18 clinically relevant configurations. The phantom was scanned at a Siemens 1.5 Tesla Aera scanner using a T1MPRAGE sequence optimized for clinical use and a T1TSE sequence optimized for research purposes. We varied slice acquisition plane and phase encoding direction and calculated the distance between the caudal tip of the DBS lead MRI artifact and the actual tip of the lead, as estimated from MRI reference markers.ResultsImaging parameters and lead configuration substantially altered the difference in the depth of the lead within its MRI artifact on the scale of several millimeters − with a difference as large as 4.99 millimeters. The actual tip of the DBS lead was found to be consistently more rostral than the tip estimated from the MR image artifact. The smallest difference between the tip of the DBS lead and the tip of the MRI artifact using the clinically relevant sequence (i.e., T1MPRAGE) was found with the sagittal acquisition plane and anterior-posterior phase encoding direction.Discussion/ConclusionThe actual tip of an implanted DBS lead is located up to several millimeters rostral to the tip of the lead’s artifact on postoperative MR images. This distance depends on the MRI sequence parameters and the DBS system’s extracranial trajectory. MRI parameters may be altered to improve this localization.

Publisher

Cold Spring Harbor Laboratory

Reference70 articles.

1. A Randomized Trial of Deep-Brain Stimulation for Parkinson’s Disease;NEJM,2009

2. Deep brain stimulation for Parkinson’s disease

3. Deep-Brain Stimulation for Essential Tremor and Other Tremor Syndromes: A Narrative Review of Current Targets and Clinical Outcomes;Brain Sci,2020

4. Long-term results of thalamic deep brain stimulation for essential tremor

5. Bilateral Deep-Brain Stimulation of the Globus Pallidus in Primary Generalized Dystonia;NEJM,2009

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3