Abstract
ABSTRACTSpinal cord neuromodulation has gained much attention for demonstrating improved motor recovery in people with spinal cord injury, motivating the development of clinically applicable technologies. Among them, transcutaneous spinal cord stimulation (tSCS) is attractive because of its non-invasive profile. Many tSCS studies employ a high-frequency (10 kHz) carrier, which has been reported to reduce stimulation discomfort. However, these claims have come under scrutiny in recent years. The purpose of this study was to determine whether high-frequency tSCS is more comfortable at therapeutic amplitudes, which evoke posterior root-muscle (PRM) reflexes. In 16 neurologically intact participants, tSCS was delivered using a 1-ms long monophasic pulse with and without a high-frequency carrier. Stimulation amplitude and pulse duration were varied and PRM reflexes were recorded from the soleus, gastrocnemius, and tibialis anterior muscles. Participants rated their discomfort during stimulation from 0-10 at PRM reflex threshold. At PRM reflex threshold, high-frequency stimulation (0.87 ± 0.2) was equally comfortable as conventional stimulation (1.03 ± 0.18) but required approximately double the charge to evoke the PRM reflex (conventional: 32.4 ± 9.2 µC; high-frequency: 62.5 ± 11.1 µC). Strength-duration curves for high-frequency stimulation had a rheobase that was 4.8X greater and a chronaxie that was 5.7X narrower than the conventional monophasic pulse, indicating that high-frequency stimulation was less efficient in recruiting neural activity in spinal roots. High-frequency tSCS is equally as comfortable as conventional stimulation at amplitudes required to stimulate spinal dorsal roots.GRAPHICAL ABSTRACT
Publisher
Cold Spring Harbor Laboratory