Abstract
AbstractThis paper proposes a tool for automatic and optimal tuning of pulse amplitude and width for sequential parameter estimation (SPE) of the membrane time constant and input–output curve in closed-loop electromyography-guided (EMG-guided) controllable transcranial magnetic stimulation (cTMS). A normalized depolarization factor is defined which separates the optimization of the pulse amplitude and width. Then, the pulse amplitude is chosen by the maximization of the Fisher information matrix (FIM), while the pulse width is chosen by the maximization of the normalized depolarization factor. The simulation results confirm satisfactory estimation. The results show that the normalized depolarization factor maximization can identify the critical pulse width, which is an important parameter in the identifiability analysis, without any prior neurophysiological or anatomical knowledge of the neural membrane.
Publisher
Cold Spring Harbor Laboratory
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献