Abstract
AbstractPurposeTo evaluate the effects of negative periocular pressure (NPP), and concomitant intraocular pressure (IOP) lowering, on the biomechanics of the optic nerve head (ONH) and cornea.MethodsWe developed a validated finite element (FE) model of the eye to compute tissue biomechanical strains induced in response to NPP delivered using the Multi-Pressure Dial (MPD) system. The model was informed by clinical measurements of IOP lowering and was based on published tissue properties. We also conducted sensitivity analyses by changing pressure loads and tissue properties.ResultsApplication of -7.9 mmHg NPP decreased strain magnitudes in the ONH by c. 50% while increasing corneal strain magnitudes by c. 25%. Comparatively, a similar increase in corneal strain was predicted to occur due to an increase in IOP of 4 mmHg. Sensitivity studies indicated that NPP lowers strain in the ONH by reducing IOP and that these effects persisted over a range of tissue stiffnesses and spatial distributions of NPP.ConclusionsNPP is predicted to considerably decrease ONH strain magnitudes. It also increases corneal strain but to an extent expected to be clinically insignificant. Thus, using NPP to lower IOP and hence decrease ONH mechanical strain is likely biomechanically beneficial for glaucoma patients.Translational RelevanceThis study provides the first description of how NPP affects ONH biomechanics and explains the underlying mechanism of ONH strain reduction. It complements current empirical knowledge about the MPD system and guides future studies of NPP as a treatment for glaucoma.
Publisher
Cold Spring Harbor Laboratory
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献