Abstract
AbstractElectrophysiological recording with a new probe often yields better signal quality than with a used probe. Why does the signal quality degrade after only a few experiments? Here, we considered silicon probes in which the contacts are densely packed, and each is coated with a conductive polymer that increases its surface area. We tested 12 Cambridge Neurotech silicon probes during 61 recording sessions from the brain of 3 marmosets. Out of the box, each probe arrived with an electrodeposited polymer coating on 64 gold contacts, and an impedance of around 50k Ohms. With repeated use, the impedance increased and there was a corresponding decrease in the number of well-isolated neurons. Imaging of the probes suggested that the reduction in signal quality was due to a gradual loss of the polymer coating. To rejuvenate the probes, we first stripped the contacts, completely removing their polymer coating, and then recoated them in a solution of 10 mM EDOT monomer with 32 uM PSS using a current density of about 3mA/cm2for 30 seconds. This recoating process not only returned probe impedance to around 50k Ohms, it also yielded significantly improved signal quality during neurophysiological recordings. Thus, insertion into the brain promoted loss of the polymer that coated the contacts of the silicon probes. This led to degradation of signal quality, but recoating rejuvenated the probes.