AbstractEpidural electrical stimulation (EES) of the spinal cord has been FDA approved and used therapeutically for decades. However, there is still not a clear understanding of the local neural substrates and consequently the mechanism of action responsible for the therapeutic effects. Epidural spinal recordings (ESR) are collected from the electrodes placed in the epidural space. ESR contains multi-modality signal components such as the evoked neural response (due to tonic or BurstDR™waveforms), evoked muscle response, stimulation artifact, and cardiac response. The tonic stimulation evoked compound action potential (ECAP) is one of the components in ESR and has been proposed recently to measure local neural activation during EES. Here, we first review and investigate the referencing strategies, as they apply to ECAP component in ESR in the domestic swine animal model. We then examine how ECAP component can be used to sense lead migration, an adverse outcome following lead placement that can reduce therapeutic efficacy. Lastly, we show and isolate concurrent activation of local back and leg muscles during EES, demonstrating that the ESR obtained from the recording contacts contain both ECAP and EMG components. These findings may further guide the implementation of recording and reference contacts in an implantable EES system and provide preliminary evidence for the utility of ECAP component in ESR to detect lead migration. We expect these results to facilitate future development of EES methodology and implementation of use of different components in ESR to improve EES therapy.