Author:
Mandal Ranajay,Babaria Nishant,Cao Jiayue,Lu Kun-Han,Liu Zhongming
Abstract
AbstractBioelectric stimulation during concurrent magnetic resonance imaging (MRI) is of interest to basic and translational studies. However, existing stimulation systems often interfere with MRI, are difficult to use or scale up, have limited efficacy, or cause safety concerns. To address these issues, we present a novel device capable of supplying current stimulation synchronized with MRI while being wirelessly powered by the MRI gradient fields. Results from testing it with phantoms and live animals in a 7 Tesla small-animal MRI system suggest that the device is able to harvest up to 72 (or 18) mW power during typical echo-planar imaging (or fast low angle shot imaging) and usable for stimulating peripheral muscle or nerve to modulate the brain or the gut, with minimal effects on MRI image quality. As a compact and standalone system, the plug-and-play device is suitable for animal research and merits further development for human applications.
Publisher
Cold Spring Harbor Laboratory
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献