EFFICACY AND CORRELATION OF INTRAOCULAR PRESSURE MEASURED BY NON-CONTACT, REBOUND AND GOLDMANN APPLANATION TONOMETER WITH CORNEAL THICKNESS, TRUE IOP AND CORNEAL CURVATURE IN A TERTIARY HOSPITAL.

Author:

Das Santanu,Nagesh Nayana,L Kiran Kumar,Shetty Sundeep

Abstract

ABSTRACT: AIM-To compare the intraocular pressure (IOP) measured by Non-Contact tonometer (NCT), Rebound tonometer (RBT) and Goldmann Applanation tonometer (GAT) and their correlation with central corneal thickness (CCT), true IOP and corneal curvature. Reliability of each tonometer. METHODS-500 random patients aged 18 years and above were taken up for the study. Patients with anterior and posterior segment pathologies like corneal ulcer, leukoma, staphyloma, corneal lacerations, ectatic corneal conditions, corneal dystrophies, oedema, perforations, acute angle closure glaucoma, retinal detachments, vitreous haemorrhage and unwilling patients were excluded from the study. IOP was recorded using NCT, RBT and GAT after assessing the patient's visual acuity. Following IOP measurement, central corneal thickness (CCT) of each patient was measured using pachymetry. Kvalues were measured using an autorefractometer. All the data were collected and tabulated for statistical analysis to obtain results. RESULTS: The mean CCT in males was 0.5350 mm and in females 0.5340 mm respectively. The mean IOP measured by NCT is 16.43 mm hg whereas the mean IOP measured by GAT is 15.43 mm hg. IOP measured by NCT is significantly higher than the IOP measured by GAT (p<0.001). When NCT and RBT are compared NCT values are significantly higher than that of RBT(p<0.001). Although the mean RBT IOP 15.83 mm hg is higher than the mean GAT IOP of 15.42 mm hg the values are not statistically significant. When correlated with CCT all the tonometers showed significant correlation with GAT showing the strongest correlation. NCT overestimates IOP in normal, thin and thicker corneas when compared to GAT and are statistically significant. RBT also overestimates in the normal and thick corneas when compared to GAT but their values are much closer to GAT values in thinner corneas. The IOP measured by all the 3 tonometers correlated with corrected IOP with NCT showing the best correlation followed by GAT. There was no significant correlation between K and IOP in our study. CONCLUSION-From the present study we can conclude that IOP measured by NCT and RBT is higher than GAT. NCT values are significantly higher than GAT values in thin and normal corneas whereas it overestimates more in thicker corneas. RBT values are significantly higher than that of GAT in normal and thick corneas. All the tonometers show significant correlation with CCT with GAT showing the strongest correlation. So, it is always advisable to measure the corrected IOP for each patient after considering the CCT. Keywords- Non-contact tonometer; Goldmann applanation tonometer; central corneal thickness; intraocular pressure; Rebound tonometer, true IOP, Corneal curvature.  

Publisher

CAB PUBLISHER

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3