Stability Study and Handling Recommendations for Multiresidue Pesticide Mixes under Diverse Storage Conditions for LC–MS/MS and GC–MS/MS

Author:

Wiest Landon A1ORCID,Hepner Jana R1ORCID,Fisher Jason E1ORCID,Risha Karen M1ORCID,Lidgett John H1ORCID,Ballarotto Valerie N1ORCID,Konschnik Joseph D1ORCID

Affiliation:

1. Restek Corporation , 110 Benner Circle , Bellefonte, PA 16823, United States

Abstract

Abstract Background In response to the growing global need for pesticide residue testing, laboratories must develop versatile analytical methods and workflows to produce scientifically sound results. One of the many challenges faced by food chemists is acquiring suitable pesticide certified reference materials (CRMs) to calibrate analytical equipment, monitor method performance, and confirm the identity and concentration of hundreds of pesticide residues in food samples. CRM producers invest considerable resources to ensure the stability of their products. Objective To present proper CRM handling and storage practices as guidance to ensure stability based on the results of several multiresidue pesticide stability studies. Methods The open ampoule and combined multiresidue mix studies were conducted under controlled conditions. New ampoules containing multiresidue pesticide CRM mixtures were opened and compared to previously opened ampoules at multiple intervals while stored under freezing and refrigerated temperatures. Both LC- and GC-amenable pesticides (>200 residues) were combined and stored under typical laboratory conditions. Studies were performed with and without celery matrix. Results The open ampoule study showed high levels of stability for all mixtures. All GC residues remained stable over the duration of the experiment. A week after opening LC multiresidue pesticide mixtures showed minor degradation. After combination of the multiresidue pesticide mixtures, degradation occurred rapidly for both the GC and LC mixtures. Conclusion Multiresidue pesticide mixtures are stable as ampullated until they are opened. Once the contents of a kit were opened and combined, decreasing stability was observed over time. This was true for both the LC and GC kits. Working mixtures of CRMs for instrument calibration should be made daily. Highlights This article shows a novel approach for measuring stability of CRM mixes. In-depth analysis of multiresidue pesticide mixtures and the stability that can be expected before and after mixing under typical storage conditions is described.

Publisher

Oxford University Press (OUP)

Subject

Pharmacology,Agronomy and Crop Science,Environmental Chemistry,Food Science,Analytical Chemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3