Development and Validation of a Method for Trace Level Zinc Quantification in Pharmaceutical Zinc Supplements Using a Carboxyl Functional Group Packed Column and Refractive Index Detector

Author:

Vanga Nagavardhana Reddy1,Kesamsetty Venkata Ratnam1,Ummiti Kumaraswamy2ORCID,Ratnakaram Venkata Nadh1,Bapatu Hanimi Reddy3ORCID

Affiliation:

1. Industrial Chemical Product Development and Analysis Centre, Department of Chemistry, GITAM School of Science, GITAM Deemed to be University-Bengaluru Campus , Nagadena Halli , Bengaluru, Karnataka 561203, India

2. Koneru Lakshmaiah Education Foundation, Department of Chemistry, College of Engineering , Green Fields, Vaddeswaram , Guntur, Andhra Pradesh 522502, India

3. JNT University, Department of Chemistry , Hyderabad, Telangana 500085, India

Abstract

Abstract Background Zinc helps with cell division, growth, wound healing, and carbohydrate breakdown. Humanbeings have to obtain zinc from food or supplements because our bodies do not produce it naturally. In view of the greater advantages (such as low cost, time of analysis, and stability-indicating) compared to other quantification methods (titration, ion chromatography, Atomic absorption spectroscopy) proposed in the literature, a refractive index detector coupled with HPLC has been used in quantification of zinc. Objective The goal of this research is to develop and validate a sensitive, low-cost, high-resolution, and stability- indicating method for detecting and quantifying zinc levels in zinc supplement pharmaceutical products (injectables, tablets, and capsules). Methods A novel isocratic reverse-phase HPLC method with a refractive index detector (with sensitivity 64, detector temperature 35°C, and positive polarity) was developed using a carboxyl functional group packed column and 0.8% (v/v) formic acid as the mobile phase to detect and quantify zinc content. Results The reported method has a good optimal sensitivity (LOQ: 0.006 mg/mL and LOD: 0.002 mg/mL). The correlation coefficient (r) obtained from the zinc calibration plot was greater than 0.998, indicating that the method was linear and that there was a strong correlation between zinc concentration (0.006 mg/mL to 0.375 mg/mL) and peak response. The accuracy at LOQ level was found to be 95–105% and 97–103% at the remaining levels (50, 100, and 150%). Conclusion The proposed method was successfully developed and validated as per International council for harmonisation (ICH) guidelines. Therefore, this method can be used for the quantitative testing of zinc in the QC laboratory. Highlights A novel method was developed for zinc levels determination in pharmaceutical products using HPLC with a refractive index detector. The present approach has a quick run time of 10 min and is inexpensive.

Publisher

Oxford University Press (OUP)

Subject

Pharmacology,Agronomy and Crop Science,Environmental Chemistry,Food Science,Analytical Chemistry

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Medical geology of zinc;Medical Geology of Africa: a Research Primer;2024

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3