Comparing Central Aortic Pressures Obtained Using a SphygmoCor Device to Pressures Obtained Using a Pressure Catheter

Author:

Esposito Cara1,Machado Priscilla1,Cohen Ira S2,Mehrotra Praveen2,Savage Michael2,Fischman David2,Davis Marguerite2,Ruggiero Nicholas2,Walinsky Paul2,McDonald Maureen E3,Dickie Kristopher4,Forsberg Flemming1,Dave Jaydev K1ORCID

Affiliation:

1. Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA

2. Division of Cardiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA

3. Department of Medical Imaging and Radiation Sciences, Thomas Jefferson University, Philadelphia, Pennsylvania, USA

4. Clarius Mobile Health, Vancouver, British Columbia, Canada

Abstract

Abstract BACKGROUND This study compared aortic pressures estimated using a SphygmoCor XCEL PWA device (ATCOR, Naperville, IL) noninvasively with aortic pressures obtained using pressure catheters during catheterization procedures and analyzed the impact of a linear-fit function on the estimated pressure values. METHODS One hundred and thirty-six patients scheduled for cardiac catheterization procedure were enrolled in IRB approved studies. Catheterization procedures were performed according to standard-of-care to acquire aortic pressure measurements. Immediately after the catheterization procedure with the pressure catheters removed, while the patients were still in the catheterization laboratory, central aortic pressures were estimated with the SphygmoCor device (using its inbuilt transfer function). The error between measured and estimated aortic pressures was evaluated using Bland–Altman analysis (n = 93). A linear-fit was performed between the measured and estimated pressures, and using the linear equation the error measurements were repeated. A bootstrap analysis was performed to test the generalizability of the linear-fit function. In a subset of cases (n = 13), central aortic pressure values were also obtained using solid-state high-fidelity catheters (Millar, Houston, TX), and the error measurements were repeated. RESULTS The magnitude of errors between the measured and estimated aortic pressures (mean errors >6.4 mm Hg; mean errors >8.0 mm Hg in the subset) were reduced to less than 1 mm Hg after using the linear-fit function derived in this study. CONCLUSIONS For the population examined in this study, the SphygmoCor data must be used with the linear-fit function to obtain aortic pressures that are comparable to the measurements obtained using pressure catheters. CLINICAL TRIALS REGISTRATION Trial Numbers NCT03243942 and NCT03245255.

Funder

American Heart Association

National Institutes of Health

Publisher

Oxford University Press (OUP)

Subject

Internal Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3