Author:
von Schenck H,Falkensson M,Lundberg B
Abstract
Abstract
Hemoglobinometry according to the International Committee of Standardization in Hematology (ICSH) suffers from imprecision related to high sample dilution and from potential errors owing to sample turbidity. We have evaluated a new instrument, "HemoCue," that measures hemoglobin at two wavelengths as azide methemoglobin, without dilution. The HemoCue method is superior to the ICSH method: by correction for turbidity, it avoids false hemoglobin readings that may arise from hyperlipemia or some large M-component of the immunoglobulin M class. We find the equipment suitable for use in outpatient units.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
155 articles.