Affiliation:
1. Department of Clinical Biochemistry, Green Lane/National Women's Hospital, Auckland, New Zealand
Abstract
Abstract
The Glucocard (Kyoto Daiichi Kagaku) blood glucose meter is designed for self-monitoring of blood glucose concentrations in capillary blood through use of an electrochemical test strip. Evaluated in this laboratory, the Glucocard had CVs of 4.6%, 6.6%, and 3.5% at blood glucose concentrations of 2.4, 4.1, and 18.9 mmol/L, respectively. The meter's response varied linearly with blood glucose concentration between 2.2 and 27.8 mmol/L. Hemolysis, urate, ascorbate, and acetaminophen interfered by > 5%. Different hematocrits, in the range 0.20-0.70, did not affect the measured glucose concentration. Comparison with glucose results measured in whole blood with a NOVA Stat Profile 5 instrument yielded the following: Glucocard = 0.898 NOVA--0.184 (r = 0.995). The main advantages of the Glucocard are its small sample volume (5 microL), wide linear range, and fully automated sample-handling steps, which reduce user-related variability.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
27 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献