Affiliation:
1. Institute für Allgemeine Pharmakologie, Medizinische Hochschule Hannover, FRG
Abstract
Abstract
A specific and sensitive assay for quantifying the immunosuppressant FK506 and its metabolites in blood and urine was developed. 32-O-Acetyl FK506 was synthesized and used as internal standard. FK506 and its metabolites were purified from the samples by solid-liquid extraction and were injected into a high-performance liquid chromatographic (HPLC) system linked to a mass spectrometer (MS) by particle-beam interface. The FK506 derivatives were separated from interfering material by use of a 100 x 4 mm C8 analytical column and water/acetonitrile or water/methanol gradient elution; they were detected by negative chemical ionization with methane as reagent gas. The limit of detection was 25 pg in a standard solution, and the limit of quantification in blood was 250 pg (extracted from 1 mL of blood). The CV was 11.3% at 5 ng, and no interferences with other drugs were found.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
46 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献