Author:
Tagari P C,Boullin D J,Davies C L
Abstract
Abstract
This simplified determination of serotonin in human plasma involves "high-performance" liquid chromatography with electrochemical detection. Deproteinated samples of platelet-poor plasma are injected directly into the chromatograph without preliminary chromatographic purification. The limit of quantitative detection of serotonin in plasma is 5 nmol/L (equivalent to 250 fmol injected). Intra- and inter-assay CVs ranged from 5.6% to 7.7% and 3.5% to 14.9%, respectively, for concentrations of serotonin from 0.02 to 10.0 mumol/L. Results agreed well with those by spectrofluorometry, both for plasma standards (r = 0.985, range 100 nmol/L to 100 mumol/L) and for pathological samples obtained from four carcinoid syndrome patients (r = 0.964). The method is also applicable to the concurrent measurement of 5-hydroxytryptophan and 5-hydroxyindoleacetic acid.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
58 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献