Author:
Arranz-Peña Maria-Luisa,Tasende-Mata Juan,Martin-Gil Francisco J
Abstract
AbstractWe report on the performance of four HDL-cholesterol assays: a homogeneous method based on polyethylene glycol-modified enzymes/α-cyclodextrin sulfate (PEGME; Kyowa); a homogeneous method based on polyanion–polymer/detergent (PPD; Daiichi); the usual precipitation method with phosphotungstic acid/MgCl2 (PTA); and an ultracentrifugation (UC) procedure. The homogeneous HDL-cholesterol assays (performed with automated analyzers) were precise and correlated well with the PTA and UC assays. The specificity and accuracy of the PEGME method were better than those of the PPD method.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
27 articles.
订阅此论文施引文献