Abstract
Abstract
We have established the optimal conditions for use of sodium borohydride as the reducing agent before the direct determination of mercury in water, urine, and blood by atomic absorption spectroscopy. We evaluated the effects of pH, temperature, and cupric sulfate concentration on the direct determination of both organic and inorganic compounds of mercury. Accurate and precise quantification of mercury requires that the pH be between 9.3 and 9.5, the reaction temperature above 25 degrees C, the reaction time longer than 1 min, and, for urine samples only, the cupric sulfate concentration 10 mumol/L. The detection limit of the method is 1 to 2 ng and the precision (CV) is 3.8% for blood and 4.0% for urine.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献