Affiliation:
1. Division of Clinical Pathology, Department of Pathology, Sinai Hospital of Baltimore, Inc., Belvedere at Greenspring Ave., Baltimore, Md. 21215
Abstract
Abstract
We have evaluated the performance of manually operated sampler pipets. Twenty-four instruments of various construction, manufacture, and delivery volume were evaluated. Pipet-delivered water and serum were repeatedly weighed under controlled environmental conditions, and precision and accuracy assessed. The pipets often failed to meet the manufacturer's stated tolerance limits. Consumers should calibrate these instruments in their own laboratories. Some of the manufacturer's recommendations for optimal performance are challenged. The manual samplers can substitute for oral pipets and permit safe, inexpensive, simple, and rapid transfer of liquids, if the user recognizes their operational characteristics and limitations in accuracy and precision.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献