Abstract
Abstract
We describe a chromatographic system involving a high-performance chemically-bonded reverse-phase column and fluorescence detection for measurement of indoles in urine. We controlled retention and selectivity by optimizing the methanol content and pH of the mobile phase. Six reference indoles were separated in less than 20 min; three 5-hydroxyindoles were eluted in less than 7 min. About 5-15 ng of aqueous solutions of these compounds can be detected. The combination of selectivity (from use of the chromatographic column) and fluorescence detection permitted analysis for five of the six indoles after a single urine-deproteinization step.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
66 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献