Accuracy and precision of methods for theophylline measurement in physicians' offices

Author:

Cook J D1,Platoff G E1,Koch T R1,Knoblock E C1

Affiliation:

1. Department of Pathology, University of Maryland School of Medicine, Baltimore 21201

Abstract

Abstract We tested the accuracy and precision of five theophylline methods intended for use in physicians' offices. The Syntex AccuLevel, Ames Seralyzer, and 3M Diagnostics TheoFAST methods were less reproducible (CVs 6.3% to 9.2%) than the Abbott Vision and Kodak DT-60 (CVs 2.2% to 3.3%). Caffeine interfered with the Vision, Seralyzer, and AccuLevel methods, and theobromine interfered with the Vision, Seralyzer, and TheoFAST methods. Only the DT-60 method was free from interference from any of the 24 compounds tested. Results by all methods correlated well with those by the HPLC comparison method (Clin Chem 1981;27:1931-3) and by the Abbott TDx method for assay of 100 serum (or, when appropriate, paired whole-blood) samples. The frequency of sample results differing from the comparison method by greater than 2.0 mg/L was as follows: TDx, 11%; Vision (serum), 12%; Vision (whole blood), 18%; DT-60, 14%; AccuLevel, 18%; Seralyzer, 25%; and TheoFAST, 31%. The Kodak DT-60 method was the most nearly accurate and precise among these physician's office methods. Some physician's office methods for theophylline analysis are not adequate to guide dosage adjustments.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 10 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3