Affiliation:
1. Division of Pediatric Endocrinology, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA
Abstract
Abstract
Determination of the plasma concentration of testosterone (T) is important in evaluating infants born with ambiguous genitalia and micropenis, and several commercially available kits provide a direct assay of T in unextracted plasma. Using plasma samples obtained from 36 subjects < 6 months old, we compared the concentration of plasma T measured by RIA after extraction and purification by column chromatography with the T concentration measured in a direct assay. When aliquots of samples were purified before RIA, the concentration of T was markedly lower than in the direct assay. In the first 3 weeks postpartum, results of the direct assay were 3.8-fold greater than those obtained after purification. This difference decreased over time, and by age 2 months there was fairly good agreement between the two methods. These data indicate that some direct assays of plasma T are inappropriate during the first 2 months postpartum.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
30 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献