Affiliation:
1. Radioimmunoassay Center, SID Applications Research, Beckman Instruments, Inc., Campus Drive at Jamboree Blvd., Irvine, Calif. 92664
Abstract
Abstract
In radioimmunoassays and radioassays for various constituents of serum, color quenching is a source of error when liquid-scintillation methods are used in samples that contain hemolyzed blood and samples from jaundiced patients. A quench correction can be made by using the External Standard Channels Ratio method and normalizing the ratios, or, alternatively, by chemical decolorization with tetramethylammonium hydroxide, hydrogen peroxide, and ascorbic acid.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
3 articles.
订阅此论文施引文献