Affiliation:
1. Department of Pathology, UCLA School of Medicine 90024-1732
Abstract
Abstract
In an effort to replace HPLC for whole-blood determination of cyclosporine (CsA), we compared HPLC with radioimmunoassay (RIA; INCSTAR, Cyclo-Trac SP assay), fluorescence polarization immunoassay (FPIA; Abbott TDx), and in-house modified enzyme-multiplied immunoassay technique (EMIT; Syva Co.). For blood samples from 200 various transplant recipients, RIA = 1.262 (HPLC) - 8.16, r = 0.983; FPIA = 1.200 (HPLC) + 19.90, r = 0.981; and EMIT = 1.038 (HPLC) + 11.28, r = 0.985. For segregation by transplant type, RIA, FPIA, and EMIT demonstrated positive biases of 27%, 12%, and 3%, respectively, for liver transplant recipients (n = 50) when compared with HPLC. Heart transplant recipients (n = 50) gave positive bias values of 23%, 14%, and 4% for RIA, FPIA, and EMIT, respectively. Adult renal transplant recipients (n = 50) demonstrated positive bias values of 30%, 31%, and 0% for RIA, FPIA, and EMIT, respectively. For pediatric renal transplant recipients (n = 50), positive biases of 40%, 31%, and 9% were obtained for RIA, FPIA, and EMIT, respectively. We conclude that the modified EMIT represents the best replacement for HPLC.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
29 articles.
订阅此论文施引文献