Affiliation:
1. Department of Chemistry, University of Florida, Gainesville, Fla. 32611
Abstract
Abstract
We measured electromotive force at 37 °C with pH, sodium glass, valinomycin, and the Simon calcium electrodes in synthetic electrolyte mixtures simulating serum, in cells both with and without liquid junction. Evidently these electrodes respond in a near-Nernstian manner to changes in concentration of the ion sensed. We suggest that such mixtures may serve as calibrating standards for ion-selective electrodes in clinical analysis. The effect of ionic strength on electrode response can be accounted for almost wholly by alterations in the activity coefficients. The lanthanum fluoride electrode shows promise as a reliable reference in cells without liquid junction.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
24 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献