Author:
Boone D J,Duncan P H,MacNeil M L,Smith B F,Houston B,Hearn T L
Abstract
Abstract
More than 300 laboratories participated in an interlaboratory survey of creatine kinase (CK, EC 2.7.3.2) determinations in which they analyzed seven lyophilized samples for total CK and CK isoenzymes and furnished information about their methodology. The samples were not necessarily intended to mimic typical patients' specimens but rather to determine the analytical ability of the laboratories to distinguish isoenzyme fraction CK-MB from CK-BB and to detect small but abnormal amounts of CK-BB. For total CK measurement, most laboratories used an NADP+ reduction method monitored at 340 nm (89%), and reported results in units per liter (U/L) (99%) at either 30 degrees C (34%) or 37 degrees C (60%). Despite the variety of analytical conditions, most laboratories (89%) correctly reported results within their normal range for all samples. The 287 laboratories that reported isoenzyme distributions in the samples used either cellulose acetate (37%) or agarose (44%) electrophoresis, ion-exchange chromatography (9%), or immunoinhibition (7%). Results from laboratories that used nonspecific CK-MB immunoinhibition techniques were biased when a significant amount of CK-BB isoenzyme was present.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献